SONNET 2 
CI and EAS Audio Processor

These support guides will help you handle and care for the SONNET 2 & SONNET 2 EAS Audio Processor.

Indicator Lights

Confirmation

SONNET 2 FineTuner Command Received

Status
FineTuner Echo command has been successfully received.

Program Change

SONNET 2 Program 1

Status
Program 1 is selected.


SONNET 2 Program 2

Status
Program 2 is selected.


SONNET 2 Program 3

Status
Program 3 is selected.


SONNET 2 Program 4

Status
Program 4 is selected.

Status

SONNET 2 Status

Status
If the optional status light has been activated: SONNET 2 is initialised and working.

Warning Patterns

SONNET 2 Empty Batteries

Status
Batteries are empty.
Solution
Change the batteries.


SONNET 2 Minimum Or Maximum Reached

Status
Either minimum or maximum volume or audio sensitivity has been reached.
Solution
Adjust volume or sensitivity if necessary.

Error Patterns

SONNET 2 Temporary Problem

Problem
Temporary error or electronic malfunction.
Solution
1. Turn processor off.
2. Turn processor back on.


SONNET 2 Setting Not Programmed

Status
The setting you selected has not yet been programmed, or there has been a program malfunction.
Solution
Select another setting.


SONNET 2 Temporay Malfuntion

Status
Temporary error or electronic malfunction.
Solution
1. Turn processor off.
2. Turn processor back on.


SONNET 2 Programming Error

Status
Programming failure, temporary error, or electronic malfunction.
Solution
1. Turn processor off.
2. Turn processor back on.

DL-Coil Indicator Light

Confirmation

DL-Coil Confirmation

Status
The DL-Coil, DL-Coil cable, and SONNET 2 are working; your SONNET 2 and implant are successfully connected (applicable to SYNCHRONY, CONCERTO, MED-EL CONCERT, SONATA, and PULSAR implants).


SONNET 2 FineTuner Command Received

Status
The DL-Coil, DL-Coil Cable, and SONNET 2 are working (applicable to C40+ and C40 implants).

No Light Signal

DL-Coil no light

Status
Your audio processor isn’t working.
Solution
1. Change the batteries. If that doesn’t solve the problem, go to 2.
2. Replace the DL-Coil cable. If that doesn’t solve the problem, contact your CI centre for assistance


Status
The indicator light has been deactivated by your audiologist.
Solution
Contact your audiologist.

Warning

DL-Coil Warning

Status
The DL-Coil and implant are disconnected.
Solution
Position the DL-Coil over the implant.


Status
SONNET 2 is in monitoring mode.
Solution
Switch SONNET 2 off and on again.


Status
The DL-Coil is positioned over the wrong implant.
Solution
Position the DL-Coil over the correct implant.


Status
The DL-Coil cable is defective. If the DL-Coil cable fails, the DL-Coil will automatically power off after 5 minutes.
Solution
1. Replace the DL-Coil cable.
2. Reposition the DL-Coil over the implant.
3. Switch your SONNET 2 off.
4. Switch your SONNET 2 back on.


DL-Coil powered off

Status
The DL-Coil has powered off.
Solution
1. Reposition the DL-Coil over the implant.
2. Switch your SONNET 2 off.
3. Switch your SONNET 2 back on.


DL-Coil cable is defective

Status
The DL-Coil cable is defective.
Solution
Replace the DL-Coil cable.

Book-Icon

Refer to your user manual for more information.